| Posted on Thursday, May 09, 2002 - 1:25 pm: |
|
Does anybody have the settings for a Sencor Mighty-Mite TC 109 Tube Tester to check some 8950 tubes???? This tester is older than the 8950 tube make,... so I am trying to find the right calibration/voltage for the tester. Thanks XLAXX |
Scrapiron63
| Posted on Thursday, May 09, 2002 - 5:26 pm: |
|
I have a Sencore TC 162, Mighty Mite VII, this is the setup for 8950 with it. Pins, 1-6-7-8-0 Filament, "B"--13 Load, "C"-- A Setup, "D"---5 Tube socket--10 If your book has the 6KV6, the 8950 will setup the same except for the voltage, 12 instead of 6, and pin 9 does not need to be on for the 8950. |
xlaxx
| Posted on Friday, May 10, 2002 - 11:46 am: |
|
Scrap- I'll try it and see if it works. I'll get back with ya.... thanks! xlaxx |
xlaxx
| Posted on Thursday, May 16, 2002 - 12:03 pm: |
|
scrap- the settings worked for the sencor t109! thanks! unfortunately, the tubes were only 50-60%.... bummer! still looking for 8950 tubes... the saga continues and the siltronix amp continues to sit gathering dust.... ain't life grand?! oh ya.... xlaxx |
|